Sensor for the measurement of peracetic acid C2H4O3, up to 1%ige sulfur- and nitric acid are tolerated.

  • Measurand(s): Peracetic acid
  • Calibration:  DIN 38409-15 “Determination of hydrogen peroxide”, ISO/DIS 7157 “Determination of hydrogen peroxide – titrimetric method”
  • Interferences:
    • ClO2 is registered with factor 1 of its measuring value
    • H2O2 is registered with factor 0.005 of its measuring value
    • O3 is registered with factor 2500 of its measuring value
  • pH range: 1 … 6
  • Pressure range: 0 … 1 bar, no pressure surges and/or fluctuations
  • Temperature range: 0 … 45 0C(not any ice crystal in test water)
  • Integrated automatic temperature compensation
  • Response time: T90 approx. 3 min
  • Flow rate: approx. 30 l/h, low flow-dependence
  • Shaft length: standard 175 mm, and up to 220 mm in length (mA-Version)
  • Connection: standard 4-pole plug, M12 male, for mA-version 2-pole terminal
  • Material: PVC-U, stainless steel 1.4571
Datasheet
Request product

Product request