Product description
- Measured variable(s): ozone
- Calibration: at the controller, via analytical determination by DPD method
Scope of supply
- OZ1.2: sensor, membrane cap, electrolyte
Areas of application
- Fresh water, must not contain surfactant
Product description
- Calibration: At the controller, via analytical determination by the DPD method (DPD-4 method), when used in seawater, the DPD-4 method (DPD-1 + DPD-3) is not selective for ozone
- Interferences:
– Cl₂ with OZ10.1H → this results in a 1.5 % increase in the reading
– Cl₂ with OZ10.1N → negligible
– ClO₂ with OZ10.1N → this results in a 6 % increase in the reading
Scope of supply
- DOSASens OZ10.1 sensor, membrane cap, electrolyte, operating instructions
Areas of application
- Tap water, demineralised water, reverse osmosis water, seawater, the membrane system is largely resistant to surfactants and mechanically robust.