Membrane-covered chlorine sensors
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Ca(ClO)₂ (calcium hypochlorite), Cl₂ (chlorine gas), chlorine generated by membrane electrolysis
- Calibration at the controller, by means of analytic determination of chlorine by DPD 1 method
Scope of supply
- CL4.2: sensor, membrane cap, electrolyte
Areas of application
- Surfactant-free fresh water with constant pH
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Ca(ClO)₂ (calcium hypochlorite), Cl₂ (chlorine gas), electrolytically generated chlorine and organic combined clorine based on iso-cyanuric acid (tested up to an iso-cyanuric acid concentration of 500 mg/l)
- In the presence of isocyanuric acid, the sensor measures the chlorine present in the isocyanuric acid as well as the chlorine that has already been released from it
Scope of supply
- CC1: sensor, membrane cap, electrolyte for use in fresh water
Areas of application
- Fresh water and seawater; surfactants are tolerated in part
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Ca(ClO)₂ (calcium hypochlorite), Cl₂ (chlorine gas), electrolytically generated chlorine
- Calibration at the controller, by means of analytic determination of chlorine by DPD 1 method
Scope of supply
- CS4: sensor, membrane cap, electrolyte
Areas of application
- Fresh water, surfactants are tolerated in part
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Ca(ClO)₂ (calcium hypochlorite), Cl₂ (chlorine gas), electrolytically generated chlorine
- Calibration at the controller, by means of analytic determination of chlorine by DPD 1 method.
Scope of supply
- CS4-...-SW: sensor, membrane cap, electrolyte
Areas of application
- Seawater at 10 μS/cm to 50 mS/cm, surfactants are tolerated to some extent
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Ca(ClO)₂ (calcium hypochlorite), Cl₂ (chlorine gas), electrolytically generated chlorine
- Calibration at the controller, by means of analytic determination of chlorine, DPD 4 method (DPD 1 + DPD 3)
Scope of supply
- CP4.0: sensor, membrane cap, electrolyte
Areas of application
- Swimming pool water, drinking water, surfactants are tolerated in part
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Ca(ClO)₂ (calcium hypochlorite), Cl₂ (chlorine gas), electrolytically generated chlorine
- Calibration at the controller, by means of analytic determination of chlorine, DPD 4 method (DPD 1 + DPD 3)
Scope of supply
- CP4.0-SW: sensor, membrane cap, electrolyte
Areas of application
- Seawater, surfactants are tolerated in part
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Cl₂ (chlorine gas), electrolytically generated chlorine
- Examination by means of analytic determination of chlorine by DPD 1 method Calibrate the sensor with chlorinated water, establish a separate measurement circuit if required
Scope of supply
- CN1.1: sensor, membrane cap, electrolyte
Areas of application
- Check for the absence of chlorine in water
Product description
- Measured variable(s): NaClO (sodium hypochlorite), Ca(ClO)₂ (calcium hypochlorite), Cl₂ (chlorine gas), chlorine generated by membrane electrolysis
- Calibration at the controller, by means of analytic determination of chlorine by DPD 1 method
Scope of supply
- DOSASens CL2.2 sensor, membrane cap, electrolyte, operating instructions
Areas of application
- Salt or seawater at concentrations of > 3.5 % up to approx. 26 % salinity, surfactant-free and at a constant pH.
Non-membrane-covered chlorine sensors
Product description
- Measured variable(s): free chlorine from sodium hypochlorite or from chlorine gas and electrolytically produced chlorine
- Calibration at the controller, by DPD 1 method
- Interferences: ozone, chlorine dioxide and chlorite are also recorded
- pH range: 5 – 9, pressure range: 0 – 8 bar
- Temperature range: 0 – 50 °C (AS2), 0 – 70 °C (AS3)
Scope of supply
- AS2, AS3: sensor, electrolyte sleeve, electrolyte
Areas of application
- Fresh water, especially drinking water, up to max. 70 °C
Product description
- Measured variable(s): free chlorine from sodium hypochlorite or from chlorine gas and electrolytically produced chlorine
- Calibration at the controller, by DPD 1 method
- Interferences: Ozone, chlorine dioxide and chlorite are also recorded
- pH range: 5 – 9
- Pressure range: 0 – 8 bar
- Temperature range: 0 – 50 °C
Scope of supply
- DOSASens ILS2, sensor, electrolyte sleeve, electrolyte, operating instructions
Areas of application
- Fresh water, especially drinking water, up to max. 50 °C